YH1102 Haze Meter can easily achieve ASTM D1003 non-compensation method, full light transmittance, haze test.Open sample bin can be vertically and horizontally tested to accommodate more samples to be tested.The YH1102 haze meter uses a PD array detector to meet the CIE V(λ)2 degree visual response. The compensation method can be used to measure the light transmittance and haze with high precision and repeatability.
Technical Advantage
1.Conforms to ASTM standards
Able to meet the test standard requirements of different users, in accordance with ASTM D1003/1044,GB/T 2410, JJF 1303-2011,CIE 15.2,JIS K7105,JIS K7361, JIS K 7136
2.Easy to operate, faster and more accurate measurement
The YH1102 haze meter is equipped with a large-size touch screen for easy operation.With a PD array detector, CIE V(λ)2 degree visual response enables high precision and repeatable transmittance and haze measurements.USB data output device for docking with laboratory system.
3.Dynamic measurement
Independent light source detector and temperature sensor, constantly monitor light source and environmental change, ensure the reliability of test data.
4.Easy to measure and widely applicable to samples
Open measuring area, vertical and horizontal testing, suitable for more samples to be tested.
5.Quality control software
It provides powerful software for measuring and analyzing haze and light transmittance, which is suitable for quality monitoring and tabulated management of haze and light transmittance data in various industries.The management of users will be digitized at the PC end, the difference of haze and light transmittance will be compared, and the test report form will be generated to facilitate customer customization and management.
Application
YH1102 haze meter is widely used in glass processing, plastic processing, film processing, screen processing, packaging industry, liquid pharmaceutical analysis and other aspects.
Technical Specification
Product Model | YH1102 Haze Meter |
Optical Geometry | Transmittance:0/D(Parallel light illumination, diffuse viewing);Conform to the standards:ASTM D1003/1044, ISO 14782,GB/T 2410, JJF 1303-2011,CIE 15.2,JIS K7105,JIS K7361,JIS K 7136 |
Characteristic | The instrument can easily achieve ASTM D1003 non-compensation method, full light transmittance, haze test. Open measuring area for vertical and horizontal testing. In glass processing, plastic processing, film, screen processing, packaging and other industries are widely used in transmittance,haze detection. |
Integrating Sphere Size | Φ154mm |
Light Source | 400nm to 700nm, combined LED light |
Spectrophotometric Mode | / |
Sensor | The PD array detector, satisfies CIE V(λ)2 degree visual response |
Wavelength Range | / |
Wavelength Interval | / |
Semiband Width | / |
Measurement range | 0~100% |
Measuring Aperture | Φ20mm/Φ8mm(Dual Aperture) |
Sample Size | Thickness <170mm |
Color Space | / |
Color Difference Formula | / |
Other Colorimetric Index | Haze(ASTM D1003/1044),Transmittance T(ASTM) |
Observer Angle | 2° |
Illuminant | D65,A,C |
Displayed Data | PASS/FAIL Result |
Measuring Time | About1.5s |
Resolution | 0.01unit |
Repeatability | Φ20mm,within 0.05(After preheating and calibration of the instrument, standard deviation values of standard haze tablets with a test haze of about 30 for 30 times were taken at intervals of 5s.) |
Inter-instrument Error | Φ20mm,with 0.4(After the instrument is preheated and calibrated, the standard deviation between the standard haze tablet and the reference value is tested at an interval of 5s.) |
Dimension | L*W*H=290X211X511mm |
Weight | About7.6kg |
Power | AC 24V, 3A Power adapter power supply |
Illuminant Life Span | 5 years, more than 3 million times measurements |
Display | 7-inch TFT color LCD, Capacitive Touch Screen |
Data Port | USB, Print serial port |
Data Storage | Standard 1000 pcs,sample 20000 pcs |
Language | Simplified Chinese,Traditional Chinese,English |
Operating Environment | 0~40℃(32~104°F) |
Storage Environment | -20~50℃(-4~122°F) |
Standard Accessory | Power Adapter, User Guide,Quality Management Software (download from the official website),USB cable,0% calibration box, measuring aperture. |
Optional Accessory | Micro-printer,Test fixture, standard haze sheet, foot switch |
Note: | The specifications are subject to change without notice. |
With 12 rigorous quality inspection processes in place, we ensure a high yield rate of up to 98% for batch processed precision parts.
FAQs About Haze Meters
High Measurement Accuracy and Stability: Low ΔE fluctuation (e.g., NH300: ΔE < 0.07; NR10QC: ΔE ≤ 0.03 short-term repeatability); Reliable sensors (CMOS dual-beam/silicon photodiode) and long-lasting light sources (1.6M–3M measurements over 5 years).
User-friendly Design and Convenient Operation: Intuitive interfaces (e.g., NH300’s “fool-proof” operation), auto-calibration, and ergonomic grips for extended use; Versatile positioning (light/cross alignment) for precise measurements.
Robust Design: Our colorimter instruments are designed to withstand rigorous use, providing long-term reliability. Rechargeable lithium batteries enable 5,000–6,000 measurements per charge (e.g., NH300/TS7030), ensuring cost-effective durability.
PC Software Support for Functionality Expansion: PC software (e.g., CQCS3) supports color difference analysis, chromaticity indexing, and sample library management.
Global customer support: Our global presence across regions ensures you get the service and maintenance support you need to keep your instrument at peak performance.
Multiple Measurement Apertures and Application Scenarios: Multiple measurement apertures, different models have different measurement apertures and some models offer multiple optional apertures. Applicable to plastic electronics, paint and ink, textile and garment printing and dyeing, printing, ceramics, automotive, food, medicine and other industries for color quality control and color difference detection.
Opaque liquids and powders can be put into a cuvette for color measurement.
Yes, we offer free samples for test, you are welcome to order samples to experience our excellent quality and service.
a. When the colorimeter is disconnected to the PC, "Connection Timeout" will occur. At the moment, you have to check whether the USB cable is well connected the colorimeter and the PC. Then you can re-plug the USB to try out.
b. After ensuring the colorimeter is connected to the PC with USB cable, turn on the colorimeter and enter the main menu. Select "Comm" and press "Enter" key to start communication.
c. Turn off CQCS3 software, and reopen it.
There are three measuring apertures for NH310: Φ8mm (standard accessory), Φ4mm (standard accessory), extended Φ8mm (optional accessory)
a. Turning on
b. Removing measuring aperture, install the needed measuring aperture.
c. Select "Settings-Aperture Setting" in main menu to select corresponding measuring aperture.
d. After selecting aperture, the instrument will display "White and Black Calibration" interface. White and black calibration must be performed.
e. Replace completed.