Color Spectrophotometer: Compact Precision for Color Analysis
The YS6020 Horizontal Format Desktop Spectrophotometer delivers lab-grade color precision for industrial applications. YS6020 Color Spectrophotometer is a high-tech device analyzing light’s spectral composition to measure color accurately. Using diffraction gratings or prisms, it splits light into wavelengths, enabling precise quantification of color parameters like hue, saturation, and brightness. YS6020 Offers reflectance and transmittance color measurement of a wide range of applications such as plastics, paints, ceramics, chemicals, etc.
Key Highlights
Reflectance and transmittance measurement, 256 Image Element Double Array CMOS Image Sensor
High accuracy: Repeatability up to ΔE*ab 0.15
Inter - instrument Error Control (ΔE * AB≤0.15)
7" TFT Capacitive Screen-touch Display
Data Storage: Standard 2000 Pcs, Sample 20000 Pcs
Ideal for color analysis within R&D and laboratory environments
YS6020 is a horizontal format spectrophotometer which is developed by 3nh independently with independent intellectual property. YS6020 has many features, like 7 inches TFT capacitive touch screen display, full illuminants, reflective d/8 and transmissive d/0 geometry (including or excluding UV). With very stable and precise color measurement, large storage and powerful PC software all makes YS6020 ideal for color analysis within R&D and laboratory environments.
Features
1, High configuration of hardware: 7 inches TFT Color Capacitive Touch-screen Display; Bluetooth 2.1; Concave Grating.
2, Double Array 256 Image Element CMOS Sensor; Long life-span stable xenon lamp.
3, With reflective and transmissive spectrum, accurate Lab value, good to calculate color formula and do precise color transmission.
4, Auto identify measuring aperture. Freely switchable between 3 measuring apertures: Φ 25.4mm/8mm/4mm. Users also can customize apertures.
5, Built-in temperature sensor to monitor and compensate the measuring temperature to ensure the measurement more precision.
6, Wavelength range 360nm – 780nm. Built-in 400nm cut off, more professional in UV measurement.
7, Independent light source detector, continuously monitor the condition of light sources to ensure the light source reliable.
8, Multiple measurement modes: Quality Management Mode, Sample Mode; Meet more users’ requirement.
9,A variety of optional accessories: Reflection sample holding tool, transmission fixture, micro 4mm aperture transmission test components, instrument inversion test components, applicable to more working conditions.
10, Big capacity data storage, for 20000 pieces test result.
11, Built-in camera locating.
12, More powerful extended functions at the PC software.
Application Industry
YS6020 benchtop spectrophotometer is used to do precise color analysis and transmission in laboratories. It can be widely applied in different industries, such as plastics, electronics, paint and ink, printing, garments, leather, paper, auto, medical, cosmetics, food, science institutes, laboratories.
| Model | YS6020(Pulsed xenon lamp) |
| Optical Geometry |
Reflect: di:8°, de:8°(diffused illumination, 8-degree viewing angle); SCI (specular component included)/SCE (specular component excluded);Include UV / excluded UV light source; Transmittance: di:0°, de:0° (diffuse illumination: 0° viewing); SCI (specular component included)/SCE (specular component excluded);Include UV / excluded UV light source; Haze(ASTM D1003); Conforms to CIE No.15,GB/T 3978,GB 2893,GB/T 18833, ISO7724/1, ASTM E1164, DIN5033 Teil7 |
| Integrating Sphere Size | Φ 154mm |
| Light Source Device | 360nm-780nm Xenon Lamp, 400nm cut-off Xenon Lamp |
| Spectrophotometric Mode | Concave Grating |
| Sensor | 256 Image Element Double Array CMOS Image Sensor |
| Wavelength Range | 360-780nm |
| Wavelength Pitch | 10nm |
| Semiband Width | 5nm |
| Reflectance Range | 0~200% |
| Measuring Aperture |
Reflective: Φ30mm/Φ25.4mm, Φ10mm/Φ8mm, Φ6mm/Φ4mm; Transmissive: Φ30mm/Φ25.4mm; Remark: 1. Automatic identification of switch caliber 2. Customized Configuration caliber and lens position |
| Specular Component |
Reflectance: SCI&SCE Transmittance: SCI&SCE |
| Color Space | CIE Lab, XYZ, Yxy, LCh, CIE LUV, Hunter LAB, Munsell, s-RGB, HunterLab, DIN, βxy |
| Color Difference Formula | ΔE*ab, ΔE*uv, ΔE*94, ΔE*cmc(2:1), ΔE*cmc(1:1), ΔE*00v, ΔE(Hunter), DIN ΔE99 |
| Colorimetric Index |
WI (ASTM E313, CIE/ISO, AATCC, Hunter), YI (ASTM D1925, ASTM 313), APHA/Hazen/Pt-Co, Saybolt MI (Metamerism Index), Staining Fastness, Color Fastness, Color Strength, Opacity, Gardner Index, 555 Index, Haze(ASTM D1003) |
| Observer Angle | 2° & 10° |
| Illuminants | D65,A,C,D50,D55,D75, F1,F2,F3,F4,F5,F6, F7,F8,F9,F10,F11,F12, CWF,DLF,TL83,TL84,TPL5,U30 |
| Displayed Data | Spectrogram/Values, Chromaticity Values, Color Difference Values/Graph, Pass/Fail Result, Color Offset |
| Measuring Time | About 2.4s (Measure SCI & SCE about 5s) |
| Repeatability |
Spectral reflectance: Φ25.4mm/SCI, Standard deviation within 0.07% (400 nm to 700 nm: within 0.06%) Chromaticity value: Φ25.4mm/SCI, Standard deviation within ΔE*ab 0.015 (When a white calibration plate is measured 30 times at 5 second intervals after white calibration) Spectral Transmittance: Φ25.4mm/SCI, Standard deviation within 0.07% (400 nm to 700 nm: within 0.07%) Chromaticity value: Φ25.4mm/SCI, Standard deviation within ΔE*ab 0.018 (When a white calibration plate is measured 30 times at 5 second intervals after white calibration) |
| Inter-instrument Error |
Φ25.4mm/SCI, Within ΔE*ab 0.15 (Average for 12 BCRA Series II color tiles) |
| Size | 370×300×200mm |
| Weight | About 9.6kg |
| Power Supply | DC 24V, 3A Power Adapter |
| Light Source Device Life | 5 years, more than 3 million times measurements. |
| Display | 7" TFT Capacitive Screen-touch Display |
| Data Port | USB & Print serial port |
| Data Storage | Standard 2000 Pcs, Sample 20000 Pcs |
| Language | Simplified Chinese, Traditional Chinese, English, |
| Working Environment | Temperature: 0~40℃; Humidity: 0~85% (No Condensation) |
| Storage Environment | Temperature: -20~50℃; Humidity: 0~85% (No Condensation) |
| Standard Accessory | Power Adapter, User Guide, CD Disk(PC Software), USB cable, Standard Calibration Board, Black Calibration Cavity, Transmission black baffle, Sample holder, 25.4 caliber, 8 caliber, 4 caliber, |
| Optional Accessory | Micro-printer, Transmissive Test Clamp Component,Micro Aperture(4mm) transmission test clamp component, Instrument inversion components, culture dish |
| Notes: The specifications are subject to change without notice. | |
FAQs About YS6020 Benchtop Spectrophotometers
The main instruments used to detect color are spectrophotometers and colorimeters (including photoelectric integrating colorimeters).
Spectrophotometer: High-precision option. It analyzes the full visible light spectrum to measure color accurately. Suitable for complex scenarios like textured surfaces, special effect colors, or batch consistency checks in industries such as paint and coatings.
Colorimeter (Photoelectric Integrating Colorimeter): Cost-effective and portable. It uses RGB filters to measure tristimulus values directly. Ideal for quick color difference detection in simple applications. Key Selection Tip Choose based on accuracy needs: use a spectrophotometer for high-precision color measurement, and a colorimeter for fast, basic color difference checks.
Yes, we offer free samples for test, you are welcome to order samples to experience our excellent quality and service.
Threenh Technology with excellent after sale service, Generally, Our Spectrophotometer instrument warranty is one years(Additional support period of 3 years). Within the warranty period, we can offer technique support and parts to parts change, necessarily, our engineers may go to your place for better support. Our instrument have professional After-sales Dept, if you have any technical problem or question with our goods welcome to contact us.
When detecting color differences, the first factors to consider when selecting a light source include its stability, directionality, lifespan, and the effectiveness of the ultimately obtained spectral curve. The illuminant of a colorimeter is a fixed bulb, such as a tungsten lamp,LED light or a long-life xenon lamp. However, for the same color sample, the results displayed by the instrument vary under different light sources. This is because different light sources cause different absorption and reflection of light on the sample, leading to differences in how both the human eye and the instrument perceive the color.
In general, the D65 light source is used in the application of coil steel inks for construction. The D65 light source is equivalent to average daylight. Most coil steel inks for construction are used outdoors, and sunlight is regarded as the standard light source in outdoor environments. For household appliance coil steel inks, due to their usage characteristics, they are mostly used indoors. Therefore, the A standard light source is adopted for color measurement of samples based on indoor lighting conditions. The A light source is a carefully specified tungsten light source. Other light sources, such as fluorescent light sources, can be used in many types of applications. For example, some textile factories use fluorescent light sources. Therefore, a reasonable light source should be selected as the mutually recognized measurement method based on actual usage conditions and user requirements. Once agreed upon by both parties, color measurement must be conducted under the same conditions. This helps reduce unnecessary systematic errors and human errors, achieving the optimal consistency in color measurement.
The 3nh high-precision spectrophotometric colorimeter adopts a combined LED light source with long lifespan and low power consumption, which includes UV (ultraviolet) and UV-excluded options. This design can meet the color difference detection needs of different users and supports the selection of multiple light source modes.
A benchtop spectrophotometer is not universally better—it outperforms portable units in precision and lab testing, while portable models win for on-site flexibility. Which one is better depends entirely on your testing scenarios, accuracy requirements and workflow.
Higher measurement accuracy & repeatability, excellent inter-instrument agreement
Supports full-spectrum analysis, more geometric optical designs
Larger testing aperture, compatible with curved, textured and bulk samples
Ideal for laboratory QC, R&D, formula matching and standard color calibration
Stable performance, less affected by external light and environment
Compact, lightweight, test anywhere anytime (production line, warehouse, on-site inspection)
Fast one-click measurement, no fixed power and lab space required
Perfect for incoming material inspection, on-line production spot check, field color comparison
Lower cost and easier to operate for frontline workers
Choose benchtop if you need:
Lab R&D, high-precision color grading, standard sample database establishment, strict quality control for cosmetics, plastics, textiles and food.Choose portable if you need:
On-site rapid detection, production line mobile sampling, outdoor and on-location color measurement.Discuss your project with one of our product specialists:
0086 755 26508999We respond to quote requests within 1 business day.
service@3nh.comSubmit a demo request for the product you're interested in.
Get a Demo